簡述實驗室氣路廠家潔凈氣路安裝注意事項

時間:2021-12-29 13:36:14 來源:本站 點擊:1307次

實驗室氣路廠家潔凈氣路安裝注意事項有:

施工要點說明:


1、所有不銹鋼管道兩端用塑料蓋密封,在進入施工現場后,安裝前,方可除去塑料蓋,為保證焊接的效果,自動焊時畢須在焊接管中吹Ar氣。

 

2、管道鋪設時,應注意平直,彎管處采用專用彎管器,不得徒手彎曲,切斷管道時,用專用切管器操作,嚴禁用鋸子鋸斷管道。管道切斷后,應用專用工具處理端口,嚴禁用普通銼刀處理。

 

3、在管道的行進路線中,每隔1.2-1.5米應設置一組管夾,如果特殊建筑物結構,應酌情考慮。

 

4 、管道穿墻及穿地板時,應設置套管,套管與管道之間的空隙,應采用不可燃的材料填充。

 

5、所有螺紋連接處應采用密封帶密封。

 

6、所有調節固定面板、所有出口點固定架(或固定面板)及所有管道上,都應貼設標有對應氣體的成分及濃度的氣體標頭。

 

7、所有系統部件安裝完畢后,應用高純氮氣進行三遍以上的大流量吹掃。

 

8、 吹掃完畢后,用高純氮氣進行檢漏保壓測試,測試壓力應為工作壓力的1.2-1.5倍。

 

9、整個施工過程中,應注意施工安全,特別是空中作業時(在梯子或腳手架上施工),應有專人固定梯子或腳手架。在土建尚未完工的工地施工時,須***佩帶安全帽,穿安全鞋。